Retour à tous les événements

Laboratory X-ray PhotoElectron Emission Microscopy

A monochromatic, focused Al Kα X-ray source (hv = 1486.6 eV) from Thermo Scientific has been installed on our Photoemission electron microscope PEEM. The X-ray source was funded by the French National Project PEPR MicroElec and the European Union’s Horizon Europe research and innovation program (Grant agreement No. 101135656) Ferro4EdgeAI.

The figure shows a survey spectrum obtained in spectroscopic mode on a single 40x40µm2 HfZrO2-based ferroelectric capacitor after careful thinning of the top TiN electrode down to 1.9 nm, allowing detection of photoemission from the underlying ferroelectric. The goal is to probe the TiN/HZO interface chemistry after standard lithography process. We also obtained in the same mode a spectrum of the Hf4f core level from the buried interface.

Figure 1: a) 3D view of the new X-ray source fitted to the PEEM analysis chamber. b) Survey spectrum obtained from the TiN/HZO structure. c) In blue, area selection on 40x40µm2 capacitor for XPEEM data acquisition. d) Hf 4f core level spectrum obtained from the buried HZO layer in the region of interest defined on c).

These preliminary results are the first reported laboratory core-level XPEEM results from a buried interface, opening up the perspective of characterizing the chemical and electronic structure of microscopic electronic devices in the laboratory.

Précédent
Précédent
26 octobre

XRD at Soleil, France